Electromagnetic Compatibility
Note: This equipment is for use in a professional healthcare environment� It is not for use in the
RF shielded room of a medical electrical system for magnetic resonance imaging, where the
intensity of EM disturbances is high�
Note: This equipment is not likely susceptible to interference from HF surgical instruments in the
Special Environment of being in close proximity to an active HF surgical instrument� In the case
that HF surgical interference is observed, adjust the separation distance of the equipment�
Warning
When this device is connected with other electrical equipment, leakage currents may
be additive� To minimize total leakage current per patient, ensure that all systems are
installed according to the requirements of IEC 60601-1�
Cautions
• Portable and mobile RF communications equipment may aect the normal function
of the display
• Do not use cables or accessories other than those provided with the display and
transmitter, as this may result in increased electromagnetic emissions or decreased
immunity to such emissions�
• If the display is used adjacent to or stacked with other equipment, observe and
verify normal operation of the display and transmitter in the conguration in which
it will be used prior to using it in a surgical procedure� Consult the tables below for
guidance in placing the display and transmitter
Like other electrical medical equipment, the VisionPro SYNK 26 " Wireless LED Display and SYNK
Wireless Transmitter require special precautions to ensure electromagnetic compatibility with
other electrical medical devices� To ensure electromagnetic compatibility (EMC), the display and
transmitter must be installed and operated according to the EMC information provided in this
manual� The display and transmitter have been designed and tested to comply with IEC 60601-1-
2 requirements for EMC with other devices�