To obtain optimal performance and prevent damage,
perform one of the following sterilization procedures:
“Flash” Autoclave:
Gravity displacement sterilizer
270-272 °F (132-134 °C)
Unwrapped in an instrument tray
10-minute minimum exposure time
Hi Vac:
Pre-vacuumed sterilizer
270-272 °F (132-134 °C)
Wrapped or unwrapped
4-minute minimum exposure time
8-minute minimum dry time
ETO:
100% ETO
120-135 °F (49-57 °C)
Wrapped in an instrument tray or fully perforated
sterilization box
2-hour 30-minute exposure time, 8-hour minimum
aeration time
250 °F Gravity:
Gravity displacement sterilizer
250-254 °F (121-123 °C)
Wrapped in an instrument tray or fully perforated
sterilization box
50-minute exposure time
8-minute minimum drying time
270 °F Gravity:
Gravity displacement sterilizer
270-272 °F (132-134 °C)
Wrapped in an instrument tray or fully perforated
sterilization box
35-minute minimum exposure time
8-minute minimum dry time
*Validation is based on the Association for the Advancement of
Medical Instrumentation (AAMI) protocol.
To Sterilize Handpieces
NOTE: After sterilization, allow the equipment to cool
to room temperature to ensure a comfortable operating
temperature.
To Sterilize Battery Packs
See the care instructions supplied with the battery
packs.
WARNINGS:
Clean and sterilize handpieces and batteries before
first and every use.
Prior to cleaning and sterilization, remove all acces-
sories from the handpieces.
Sterilization Recommendations*