6206-001-700 Rev- 2006/05
www.stryker.com 13
Guidance and manufacturer's declaration - electromagnetic immunity
The System 6 handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of the
System 6 handpiece should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
n/a
n/a
Portable and mobile RF communications equipment
should be used no closer to any part of the System 6
handpiece, including cables, than the recommended sep-
aration distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d=1.67√P
d=1.67√P
80 MHz to 800 MHz
d=2.33√P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m)
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Specifications (cont’d)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker System 6 6206 and is the answer not in the manual?

Stryker System 6 6206 Specifications

General IconGeneral
BrandStryker
ModelSystem 6 6206
CategoryMedical Equipment
LanguageEnglish

Related product manuals