Guidance and manufacturer's declaration - electromagnetic immunity
The System 6 handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of the
System 6 handpiece should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±2, 4, 6 kV contact
±2, 4, 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
n/a
n/a
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
n/a
n/a
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0,5 cycle
n/a
40% U
T
(60% dip in U
T
)
for 5 cycles
n/a
70% U
T
(30% dip in U
T
)
for 25 cycles
n/a
<5% U
T
(>95% dip in U
T
)
for 5 sec
n/a
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristics of a
typical location in a typical commercial
or hospital environment.
NOTE: U
T
is the alternating current mains voltage prior to application of the test level.
Specifications (cont’d)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker System 6 6206 and is the answer not in the manual?

Stryker System 6 6206 Specifications

General IconGeneral
BrandStryker
ModelSystem 6 6206
CategoryMedical Equipment
LanguageEnglish

Related product manuals