Preventive maintenance
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by qualified personnel.
WARNING
Risk of electric shock. Disconnect power before servicing or cleaning the T/Pump.
Note: Change the water monthly or more often depending on use.
Inspect the following items:
Line cord, the plug, and the housing for cuts or cracked insulation and quality of the strain relief, if damaged
replace
Membrane panel for cuts or cracks, if damaged replace
Plastic parts on the pump for cracks
Connectors for cracks and missing O-rings, replace connectors if necessary
Hose connections and connectors function properly
Pads and hoses are free from tears, cracks, and water leaks
Refer to the T/Pump Maintenance Manual Functional Checks section that and complete during preventive maintenance.
Inspection Value
Okay
Action Needed Action Taken
Measure current
leakage <100 µA
Product serial number:
Completed by:
Date:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker T/Pump Series and is the answer not in the manual?

Stryker T/Pump Series Specifications

General IconGeneral
BrandStryker
ModelT/Pump Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals