Introduction
Contraindications
Stryker promotes the assessment of each patient and appropriate usage by the operator. Contraindications for this
product are for patients with the conditions specified below.
Heating contraindications include:
Application to a body surface with compromised blood flow (Ischemia, area under pressure, arterial insufficiency)
Application to a patient with an increased tendency to bleed (aggravates potential for hemorrhage)
Application to a body surface with possible malignancy (tissue metabolism is increased and therefore, the growth
potential of the malignant tissues)
Treatment of hematoma within the first 24 to 48 hours (potential for re-bleeding and hemorrhage). Recent sprain or
fracture (acute inflammatory response).
In combination with topical solutions whose toxicity may be affected by the application of heat
In combination with other heat sources
Cooling contraindications include:
Application to a body surface with compromised blood flow (Ischemia, area under pressure, arterial insufficiency)
Application to body surface with known vascular impairment such as frostbite, arteriosclerosis or ischemia
Application to body surface in patients with hypersensitivity to cold, such as people with Raynaud’s phenomenon,
cold urticaria, cryoglobulinemia, and paraxysmal cold hemoglobinuria
Application to body surface in patients with impaired sensation
In combination with topical solutions whose toxicity may be affected by the application of cold
Specifications
Models
TP700 / TP700C
Temperature setpoints Fahrenheit 107° F, 100° F, 95° F, 50° F
Temperature setpoints Celsius 42° C, 38° C, 35° C, 10° C
Therapy duration 20 or 30 minutes or continuous cycle
Classification Class I equipment with Type BF applied part suitable for continuous
operation. Not classified for protection against ingress of liquid. Not
classified for use in the presence of flammable anesthetics.
Size
11.5 in. x 8 in. x 8 in. 29.2 cm x 20.3 cm x 20.3 cm
Weight Reservoir empty 6.5 lb 2.9 kg
Reservoir filled with water to the
heating level
9 lb 4.0 kg
Maximum reservoir capacity 93 oz 2.75 l
Flow rate, minimum with pad attached 9 gph 34 lph
Average temperature accuracy ±2° F at 107° F
±1° C to 42° C
Maximum contact surface temperature
107° F
42° C
High limit safety temperature
110° F to 120° F
43.3° C to 49° C
Power cord International (harmonized) 3 wire cord set
Dual hose
10 ft
305 cm
Current leakage, maximum
100 µA
Ground resistance, maximum 0.5
Electrical requirements Voltage
120 VAC ±10%
Frequency: 60 Hz
Current in amperes
3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker T/Pump Series and is the answer not in the manual?

Stryker T/Pump Series Specifications

General IconGeneral
BrandStryker
ModelT/Pump Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals