80-0067-00-MO-RevF 2017-10-03 SunTech CT40 User Manual | 47
Meets or exceeds ANSI/AAMI/ISO 81060-2:2013 standard for noninvasive accuracy (±5mmHg
mean error with 8mmHg standard deviation).
Notes on Blood Pressure Data
Any blood pressure reading can be affected by the measurement site, the position of the patient, exercise, or the
patient’s physiologic condition. Environmental or operational factors which can affect the performance of the device
and/or its blood pressure reading are pacemakers and common arrhythmias such as atrial or ventricular premature
beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-clampsia, renal diseases,
patient motion, trembling, and shivering.
SP0
2
Sensor Specifications
Saturation Normal (no motion)
Pulse Heart rate w/motion
Wavelength & Output Power
RED: 660nm @ 3.5-4.5mW
IR: 905nm @ 3.5-4.5mW
RED: 660nm @ ≤ 15mW
IR: 905nm @ ≤ 15mW
For more detailed information, see the instructions for use included with each device.
Temperature Sensor Specifications
Covidien
®
Electronic Thermometer
30.1 to 42.9 °C (86.2 – 109.2 °F)
Quick Mode (Oral): ± 0.3°C
Std. Mode (Axillary/Rectal): ± 0.1°C
Direct Mode: ± 0.1°C
3-10 sec (Oral Quick Mode)
6-10 (Oral Standard Mode)
8-12 sec.(Axillary Mode)
10-14 sec.(Rectal Mode)
60-120 sec.(Direct Mode all sites)