80-0067-00-MO-RevF 2017-10-03 SunTech CT40 User Manual | 40
Power off and repower on the CT40.
Return for Service.
Run only from A/C power source.
Have battery service performed at earliest convenience.
Power off and repower on the CT40.
If repower attempt does not work, plug in power source and power off and repower on
CT40 again.
Retry reading. If issue persists, call Customer Service.
Clear the error and power off and repower on the CT40.
Replace battery.
Check and insure proper barcode and Patient ID was scanned
Rescan Patient ID
Disconnect / Reconnect cable from USB port on CT40.
Replace Barcode Scanner
Check cables. Reset Printer. Replace Printer.
Out of Range Measurements
The SunTech CT40 can measure BP that falls within the following ranges.
Systolic: Adult 40-260 mmHg, Pediatric 40-230 mmHg
Diastolic: Adult 20-200 mmHg, Pediatric 20-160 mmHg
Heart Rate: 30 - 220 BPM
The SunTech CT40 will display temperature and SpO
2
within the following ranges.
FILAC
®
Temperature: 30.1 to 42.9 °C (86.2 – 109.2 °F)
Touchless IR Temperature: 34 to 42.5 °C (93.2 – 108.5 °F)
SpO
2
: 70 – 100%