80-0067-00-MO-RevF 2017-10-03 SunTech CT40 User Manual | 38
CT40 Barcode Scanner (w/USB) with Scanner Mount
CT40/50 WiFi Dual Band USB Dongle
12. Status Messages & Alarms
If the SunTech CT40 has a problem taking a measurement, you may see a code on the display screen.
The icon will blink with the respective measurement display result;
(i.e blood pressure, SpO
2
, Temperature), while displaying the code.
Below are common error codes. Take action as directed on the screen, or as suggested in the table below.
Status Messages
Check the cuff is in the correct position.
Check that the cuff is properly tightened and tubing properly connected.
Check that there is no excessive clothing between the arm and cuff.
Check that the cuff applied is the correct size.
The patient may have been moving too much.
Take another BP reading.
No action possible. Values are outside the reportable range.
Check that the hose has no sharp bends and is not pinched.
Check that the patient is not lying on the cuff.
Check that the cuff is in the correct position.
Take another BP reading.
If cancelled reading was unintended, allow cuff to deflate & restart reading.
If cancelled reading was intended, no action needed.
Power off and repower on the CT40.