11 Calibration
The dew point calibration is not necessary when you receive a new
instrument. The new instrument is calibrated ex work and measures
accurately. The user dew point calibration should only be performed by
professionals. Please contact your service before performing a user
calibration.
The exact calibration date is printed on the certificate which is supplied
together with the sensor. The accuracy of the sensor is regulated by the
on site conditions, parameters like oil, high humidity or other impurities
can affect the calibration and furthermore the accuracy. However we
recommend to calibrate the instrument at least once per year. The
calibration is excluded from the instruments warranty. For this please
contact the manufacturer.
If the instrument has a slight drift due to ageing, temperature or other
effects a dew point and pressure calibration can be performed.