Calibration
11.1 Dew point calibration
Performing a dew point calibration is critical and the following
considerations should be taken:
• Perform dew point calibration at the working point. Foe example if
you measure around -40°C Td do the calibration at that point.
• Do not perform calibration at too high dew points, as it will cause
big errors at low dew points!
• We recommend calibration between -40°C Td and -55°C Td.
• Use highly precise reference measuring instruments.
• Maintain conditioning time of about 1 hour minimum.
Please use the calibration function and select “dew point calibration”.
Follow the steps on the screen.
If there is any doubt about correct calibration process and result, the
calibration can be removes at any time. Please follow the instruction on
the screen.