Operator‘s Guide / iMRS
/ Page 5
© Swiss Bionic Soluons Schweiz GmbH / 8834 Schindellegi / 2016
Low pulsang electromagnec elds (LPEMF) do not create electrosmog!
Even at the highest intensity sengs, the average peak eld strength of the Magnec-Reso-
nance-Smulaon System is less than or equal to 300 µT (microTesla). This is less than the eld
strengths produced by common domesc appliances and well below the recommended limit of
400 µT (microTesla) for compliance with DIN 0848.
Safety is ensured by the fact that the frequencies and eld strengths employed are on the same
magnitude as naturally-occurring electromagnec elds within the Earth’s ionosphere.
Instrucons on electromagnec compability (EMC)
Medical, electric devices are subject to special precauons in terms of EMC and must be ins-
talled and brought into operaon in accordance with the EMC instrucons in the accompanying
documents. Electromedical devices can be parcularly aected by portable and mobile RF com-
municaons equipment, e.g. mobile phones or walkie-talkies.
The manufacturer only guarantees the compliance of the device with the EMC requirement
when used with the accessories listed in 16.2. The use of other accessories can lead to an incre-
ased emission of electromagnec interference or to a reduced resistance against such interfe-
rence. The accessories listed may only be used together with an iMRS system from Swiss Bionic
Soluons Schweiz GmbH.
The device may not be operated in combinaon with other devices, nor be stacked with or loca-
ted in the proximity of these. Should such an arrangement, however, be necessary, the device
must be kept under observaon to check its operaon for the use intended when in such an
arrangement.
The propagaon of the magnec eld from the whole-body / pad applicators has largely abated
at a distance of about 1.5 metres. Persons not receiving treatment should remain outside such a
range during the course of an applicaon.
We are legally obliged, in accordance with the EMC regulaons for medical products, to
provide you with the following informaon:
Guidelines and manufacturer‘s declaraon - electromagnec interference emissions
The iMRS is intended for operaon in an ELECTROMAGNETIC ENVIRONMENT as given below.
The customer or user of the iMRS should ensure that it is operated in such an environment.