SYNRAD OEM v40 Operator’s Manual Version 4
Troubleshooting
Possible Causes:
Water-cooled lasers – Cooling water temperature is above 22 °C (72 °F) or there is inadequate water
ow through the laser.
Check that your chiller is maintaining a water temperature between 18 °C–22 °C (64 °F–72 °F) at a ow
rate of one gallon per minute (GPM). If water temperature is OK, check the ow rate. The simplest way
to do this, if a ow meter is not available, is to disconnect the cooling tubing from the chiller inlet (or the
drain) and run the cooling water for one minute into a ve-gallon bucket; you should have close to one
gallon. If there is much less than one gallon, check the cooling path for kinked or pinched cooling tubes or
check the chiller for a clogged or dirty lter.
Fan-cooled lasers – Cooling fans are not providing adequate airow to the laser.
Check that all cooling fans are clear of debris and are providing an airow rate of at least 140 CFM per fan
or at least 300 CFM if rear cooling the laser. To allow for proper airow, cooling fans must have at least
2.25" (57.2 mm) of unobstructed clearance between the outside edge of the cooling fan housing and any
mounting surface or enclosure.
Symptom:
A shutter closed condition is indicated:
INT LED – Green Interlock Open – Open
TMP LED – Green Over Temperature Open
RDY LED – Yellow Laser Ready – Closed
SHT LED – O Shutter Open – Open
LASE LED – O Laser Active – Open
Possible Causes:
The Shutter Switch (if equipped) is Closed.
If your Firestar laser is equipped with a Shutter Switch, slide the switch to the Open position. The SHT
indicator will illuminate blue.
No voltage applied to Pin 10 (Shutter Open Request) on the User I/O connector.
Check to see that a positive or negative voltage (±5–24 VDC) is applied to Pin 10, Shutter Open Re-
quest, with respect to Pin 11, Input Common, on the User I/O connector (refer to User I/O connections in
the Technical Reference” chapter for details). If your system does not provide a Shutter Open Request
signal, connect the factory-supplied Quick Start Plug to the User I/O connector on the laser’s rear panel
or wire your male DB-15 connector so that Pin 11 (Input Common) is jumpered to Pin 12 (Auxiliary DC
Power Ground) and Pin 10 (Shutter Open Request) is jumpered to Pin 4 (+5 VDC Auxiliary Power).