technical reference
80
SYNRAD OEM v40 Operator’s Manual Version 4
v40 general specications
Parameter Firestar v40
Environmental Specications
Operating Temperature
7
....................................15 °C–40 °C
Humidity ...................................................................0–95%, non-condensing
Physical Specications
v40, water-cooled
Length .......................................................................25.10 in (63.8 cm)
Width ................................................................. 5.62 in (14.3 cm)
Height ............................................................... 6.21 in (15.8 cm)
Weight ..............................................................24.0 lbs (10.9 kg)
v40, fan-cooled
Length ...............................................................24.45 in (62.1 cm)
Width (incl. fans)............................................. 8.25 in (20.9 cm)
Height ............................................................... 6.21 in (15.8 cm)
Weight ..............................................................26.0 lbs (11.8 kg)
v40, air-cooled
Length ...............................................................24.45 in (62.1 cm)
Width ................................................................. 6.22 in (15.8 cm)
Height ............................................................... 5.76 in (14.6 cm)
Weight ..............................................................23.0 lbs (10.4 kg)
* Specications subject to change without notice.
7 Published specications guaranteed at a cooling temperature of 22 °C. For air-cooled lasers, some performance degradation may
occur in ambient temperatures above 22 °C. Typically output laser power decreases 0.5–1 Watt per degree Celsius increase in
ambient temperature.