Instruction Manual
Tek-Flux 1400A
www.tek-trol.com
23
The Full-Scale Current is Correct
Select “Analog Range” to enter. Adjust the converter parameter and make sure the
ampere meter is 20 mA (±0.004 mA). Adjust the current between zero and full range,
the current function of the converter should reach exactness. The line degree of the
current output of conversion should be controlled within the scope of 0.1%.
Current Line Degree Checking
You can place the standard signal source at 75%, 50%, 25%, and check the line degree
of current output.
4.6.3 Transmitter Connection of Current Output