EasyManuals Logo

Tektronix 2465B Service Manual

Tektronix 2465B
536 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #177 background imageLoading...
Page #177 background image
Table 4-11
Parametric Measurement Period Checks
Time Mark
Setting
2
ns
5
ns
10
ns
20
ns
50
ns
100
ns
200
ns
500
ns
1
|as
2
|as
5 |as
10 |as
Minimum
Period
1.48
ns
4.45
ns
9.41
ns
19.32
ns
49.05
ns
98.6
ns
197.7
ns
495.0
ns
990.5
ns
1.981
|as
4.954
|as
9.909
|as
Maximum
Period
2.52
ns
5.55
ns
10.59
ns
20.68
ns
50.95
ns
a
101.4 ns
202.3
ns
505.0
ns
1.010 |as
2.019 |as
5.046
|as
10.09 |as
Time Mark
Setting
20 |as
50 |as
100 |as
200 |as
500 |as
1 ms
2 ms
5 ms
10 ms
20 ms
50 ms
100 ms
Minimum
Period
19.82 |as
49.55
|as
99.10 |as
198.2 |as
495.5
|as
991.0 |as
1.982 ms
4.955
ms
9.910 ms
19.82 ms
49.55
ms
99.10 ms
a
lf the 50 ns setting is not within the limits given,
b
For this setting, change MINFREQ to 10 Hz.
perform step 2 (50 ns Timing Accuracy Verification).
Maximum
Period
20.18 |_IS
50.45 |as
100.9 |as
201.8 |as
504.5 |as
1.009 ms
2.018 ms
5.045
ms
10.09 ms
20.18 ms
50.45 ms
b
100.9 ms
b
REV OCT 1996
4-27

Table of Contents

Other manuals for Tektronix 2465B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2465B and is the answer not in the manual?

Tektronix 2465B Specifications

General IconGeneral
Bandwidth400 MHz
Channels4
Vertical Sensitivity2 mV/div to 5 V/div
Maximum Input Voltage400 V (DC + peak AC)
DisplayCRT
Trigger ModesAuto, Normal, Single
Time Base Range500 ps/div to 0.5 s/div
Rise Time875 ps
Power Supply48 to 440 Hz
Dimensions495 mm

Related product manuals