Table 4-12
Parametric Measurement Volts Checks
Calibration
Generator
Setting
20 mV
50 mV
0.1 V
0.2 V
0.5 V
1 V
2 V
5 V
10 V
20 V
50 V
Min
a
PK-PK
14 mV
43 mV
90 mV
185 mV
470 mV
0.945
V
1.89 V
4.74 V
9.49 V
19.0 V
47.5 V
Max
8
PK-PK
26 mV
57 mV
110 mV
215 mV
530 mV
1.055 V
2.10 V
5.25 V
10.50 V
21.0 V
52.5 V
AVG
B
Within ±(5% of DM501A reading + 5.6 mV)
Within ±(5% of DM501A reading + 5.6 mV)
Within ±(5% of DM501A reading + 5.6 mV)
Within ±(5% of DM501A reading + 6.5 mV)
Within ±(5% of DM501A reading + 6.5 mV)
Within ±(5% of DM501A reading + 6.5 mV)
Within ±(5% of DM501A reading + 15 mV)
Within ±(5% of DM501A reading + 15 mV)
Within ±(5% of DM501A reading + 15 mV)
Within ±(5% of DM501A reading + 100 mV)
Within ±(5% of DM501A reading + 100 mV)
â– Disconnect DMM prior to selecting VOLTS measurement.
REV OCT 1988
4-28a

Table of Contents

Other manuals for Tektronix 2465B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2465B and is the answer not in the manual?

Tektronix 2465B Specifications

General IconGeneral
Bandwidth400 MHz
Channels4
Vertical Sensitivity2 mV/div to 5 V/div
Maximum Input Voltage400 V (DC + peak AC)
DisplayCRT
Trigger ModesAuto, Normal, Single
Time Base Range500 ps/div to 0.5 s/div
Rise Time875 ps
Power Supply48 to 440 Hz
Dimensions495 mm