Page #475 background image
NO
^
TROUBLESHOOT
TRIGGER SYSTEM
PERFORM
CAL 01
ANDCAL09
YES
VERIFY
A
SWEEP GATE,
B SWEEP GATE,
A
DELAY
GATE,
B
DELAY GATE,
AND
A
HOLDOFF
ARE
ALL REACHING
MEASUREMENT
PAL
WITH CORRECT LEVELS
(^
DONE
^
2465B/2467B Service
PARAMETRIC MEASUREMENT TROUBLESHOOTING
-CO
2>
oao
c>
on
d m
z>
OCA
PERFORM MEASUREMENT
THAT
IS AT
FAULT
AND
MONITOR CLOCK, RESET,
DATA, AND Q OUTPUT
OF
U980A DURING MEASUREMENT
REPLACE
U975
PERFORM MEASUREMENT
AND MONITOR Q
OUTPUT OF U980B
NO
THE PROBLEM
IS
NOT PARAMETRIC
MEASUREMENT: CHECK
FOR OPEN
CIRCUITS,
ETC.
MONITOR CNTL1
AND
CNTL2 WHILE REPEATEDLY
PRESSING ANY FRONT
PANEL SWITCH
THE PROBLEM
IS
NOT PARAMETRIC
MEASUREMENT: CHECK
FOR OPEN
CIRCUITS,
ETC.