Operating requirements
Use the oscilloscope within the required operating temperature, power, altitude, and signal input voltage ranges to provide the
most-accurate measurements and safe instrument operation.
Environment requirements
Characteristic Description
Operating temperature 0 °C to +50 °C (+32 °F to +122 °F)
For proper cooling, keep the sides and rear of the instrument clear of obstructions for
2 inches (51 mm).
Operating humidity 5% to 90% relative humidity (% RH) up to +40 °C (+104 °F)
5% to 55% RH above +40 °C up to +50 °C (+104 °F to +122 °F), Noncondensing.
Operating altitude Up to 3000 meters (9842 feet)
Power requirements
Characteristic Description
Power source voltage 100 V - 240 V
AC RMS
, ±10%, single phase
Power source frequency 50/60 Hz, 100-240 V (90-264 V)
400 Hz, 115 V (103-127 V)
Power consumption All models: 400 W maximum
Installing your instrument
2 MSO54, MSO56, MSO58 Installation and Safety Manual

Table of Contents

Other manuals for Tektronix 5 Series MSO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 5 Series MSO and is the answer not in the manual?

Tektronix 5 Series MSO Specifications

General IconGeneral
BrandTektronix
Model5 Series MSO
CategoryTest Equipment
LanguageEnglish

Related product manuals