Calibration-7704A
Service
b
.
Set
t
h
e
P
O
WER
switc
h
to
off
and
discon
n
ect
all
test
TA
BLE
4-2
eq
uipment
.
P
ower
Su
p
ply
Tolerance
ε
.
Install
t
he
p
ower
unit
in
t
he
instr
u
ment
an
d
return
t
h
e
P
O
WER
switc
h
to
on
.
d
.
Connect
t
h
e
p
recision
DC
voltmeter
to
P
in
1
of
Ρ
43
on
rear
of
Α
43,
H
orizo
n
tal
Out
p
ut
boar
d
;
see
F
ig
.
6-29
.
e
.
C
HE
C
K-M
eter
rea
d
ing
+155
volts
±5
volts
.
f
.
Disconnect
all
test
eq
ui
p
ment
.
P
ower
Su
p
ply
Output
V
oltage
Tolerance
Test
P
oint
Location
(see
F
ig
.
6-28)
-50
V
olt
±0
.20
volt
P
in
6-
P
32G
-15
V
olt
±0
.15
volt
P
in
1-
P
32G
+5
V
olt
±0
.10
volt
P
in
2-
P
32G
+15
V
olt
±0
.15
volt
P
in
4-
P
32G
+50
V
olt
±0
.50
volt
P
i
n
5-
P
32G
Control
Ι
Ilum
(+5
V
L
ig
h
ts)
+0
.2
to
-0
.5
volt
P
in
9-
P
32G