Performance ver
ication
Performance v
erication
NOTE. The performance verication procedure is not a calibration procedure. The performance verication procedure
only veries that your instrument meets key specications. For your instrument to be calibrated, it must be returned to
a Tektronix se
rvice facility.
Prerequisi
tes
Thetestsinthissectionmakeupaconrmation of performance and functionality when the following requirements are met:
The SignalVu-PC application must be version 3.11 .0x or greater.
Operate the instrument in an environment that meets the temperature, altitude, and humidity characteristics listed in
the specications.
The instrument must be completely assembled and covers installed per factory specication.
The instrument must have been operating for a warm-up period of at least for 30 minutes (after being connected to the
PC, starting the SignalVu-PC application, and SignalVu-PC is connected to the RSA306B instrument and continuously
acquiring data).
NOTE. The RSA306B does not fully power on until SignalVu-PC has established communication with the RSA306B
and is acquiring data.
The instrument must have had its last alignment routine done after at least a 30 minute warm–up period at an ambient
temperature not more than ±2 °C different than the current ambient temperature.
12 RSA306B Specications and Performance Verication

Other manuals for Tektronix RSA306B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306B and is the answer not in the manual?

Tektronix RSA306B Specifications

General IconGeneral
BrandTektronix
ModelRSA306B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals