Performance ver
ication
Test record
Print out the following test record pages and use them to record the performance test results for the RSA306B spectrum
analyzer.
Table 24: Test record: RSA306B
Instrument Serial Number:
Temperature:
Date of Calibration:
Certicate Number:
RH %:
Technician:
Frequency accuracy
Measured Frequency
(1 GHz CF)
Instability measured,
ppm [(Marker
Frequency
1E9)/1000]
Total specied
Instability (aging
plus other drift) Pass/Fail
External reference
Phaselock to 10 MHz,
0 dBm Pass/Fail
Amplitude accuracy
Reference level Error, dB Specication, dB Pass/Fail
+20 dBm < 3 GHz
±1.2 dB
+20 dBm > 3 GHz
±1.65 dB
+0 dBm < 3 GHz
±1.2 dB
+0 dBm > 3 GHz
±1.65 dB
–13 dBm < 3 GHz
±1.2 dB
–13 dBm > 3 GHz
±1.65 dB
–30 dBm < 3 GHz
±1.2 dB
–30 dBm > 3 GHz
±1.65 dB
Channel amplitude atness
Center Frequency Peak Error, dB Specication, dB Pass/Fail
22 MHz < ±1.2 dB
1280 MHz < ±1 dB
2160 MHz < ±1 dB
RSA306B Specications and Performance Verication 53

Other manuals for Tektronix RSA306B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306B and is the answer not in the manual?

Tektronix RSA306B Specifications

General IconGeneral
BrandTektronix
ModelRSA306B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals