This section lists the EMC (electromagnetic compliance), s afety, and environmental standards with which the instrument
complies.
EMC Compliance
EC Declaration of Conformity EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following
specications as listed in the Ofcial Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment for measurement, control, and
laboratory use.
1234
CISPR 11:2003. Radiated and conducted emissions, G roup 1, Class A
IEC 61000-4-2:20
01. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic eld immunity
5
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line s urge immunity
IEC 61000-4-6:2003. Conducted RF immunity
6
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
7
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European Contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standar
d may occur when this equipment is connected to a test object.
3
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
At the IEC 61000-4-11 Voltage-Interruption Transient Immunity test at 0% for 250 cycles, the EUT reboots and can take greater
than 10 seconds to recover.
5
Performance degradation information for the IEC 61000-4-3 test: Residual spurious signals can typically increase to -55 dBm with
exposure to the disturbance levels of this test.
6
Performance degradation inf
ormation for the IEC 61000-4-6 test: Residual spurious signals can typically increase to the worse of
either 55 dB below reference level or -75 dBm, with exposure to the disturbance levels of this test.
7
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test l evels