Installation
Operating Considerations
Power supply requirements
Source voltage and frequency Power consumption
100 240 V
RMS
, 50/60 Hz
115 V
RMS
, 400
Hz
450 W maximum (all models)
Environmen
tal requirements
Characteristic Description
RSA6100A Series Operating temperature +5 °C to +50 °C (+5 °C to +40 °C; when accessing the DVD drive or optional
removable hard-disk drive)
RSA5100A Series Operating temperature +10 °C to +40 °C
Non-operating temperature
-2Cto+6C
RSA6100A Humidity 90% relative humidity at 30 °C (non-condensing)
80% maxi
mum when accessing the DVD drive
Maximum w et-bulb temperature 29 °C
RSA5100A Humidity 90% up to 30 °C (non-condensing)
80% relative humidity m aximum when accessing the DVD drive
Above 3
0 °C it is limited by maximum wet-bulb temperature 29 °C
Altitude:
Opera
ting
Non-operating
Up to 3
000 m (9843 ft)
12190 m (40,000 ft)
Cooling clearance
Bottom
Side
s and back
20 mm (0.79 in)
50 mm (1.97 in)
RSA6100A Series & RSA5100A Series Q uick Start User Manual 7