Installation
Inspecting the Instrument
Run the diagnostics application (Tools > Diagnostics). If failures occur, perform the following steps to get more information
about them. You can also use the following steps as a detailed incoming inspection to verify the functionality of your
instrument. If you want to check the accuracy speci cations of your instrument, see the RSA6100A Series Real-Time
Spectrum Analyzers Specications and Performance Verication Technical Reference manual PDF (Tektronix part number
077-0251-XX) or RSA5100A Series Real-Time Signal Analyzers Specications and Performance Verication Technical
Reference manual PDF (Tektronix part number 077-0520-XX) located on the Product Documentation CD.
1. Select Tools > Diagnostics.
2. On the Diagnostics tab, click All
Modules, All Tests.
3. Click Single Test.
4. Click RUN.
The instr
ument will run through
the tests one at a time. A check
mark or X icon will appear to the
right of
each check box as that
test is completed. The check icon
means the test w as passed. The
X means
the test was failed.
5. For tes
ts that require manual
intervention, follow the on-screen
instructions to complete the tests .
NOTE. Select the Diagnostics Failure Information tab to see basic diagnostic failure information. Use the Windows Event
Viewer (a subset of the Administrative tools in the Windows Control Panel) to view failure history and nondiagnostic failures
reported by the application.
14 RSA6100A Series & RS A5100A Series Quick Start User Manual