Installation
Installation
Unpacking the Instrument
Unpack the instrument and check that you received all items listed as Standard Accessories. C heck the Te ktronix Web site
(www.tektronix.com) for the most current information, recommended accessories, instrument options, and upgrades.
Standard Accessories
Accessory
Tektronix part
number
FCA3000 and FCA3100 Series Timer/Counter/Analyzer, MCA3000 Series Microwave
Counter/Analyzer Quick Start User Manual (English/S implified Chinese/Japanese)
071-2787-xx
Product documentation CD-ROM
063-4288-xx
NI LabV IEW SignalExpress CD-ROM
063-4253-xx
Power cord
Country-specific
Type-N male to BNC adapter (FCA3003, FCA3020, FCA3103, and FCA3120 only) N/A
Certificate of Calibration N/A
Operating Considerations
Pleas
e verify that the area in which you will use the instrument meets the following operating requirements:
Requirement Description
Clearance Top and bottom: 0 mm (0 in)
Left
and right side: 51 mm (2 in)
Rear: 51 mm (2 in)
Lin
ePower
100 – 240 VAC ±10%, CAT II: 50 – 400 Hz ±10%: 40 W
Tem
perature
Operating: 0 °C to +50 °C (32 °F to 122 °F)
Nonoperating: –40 °C to +71 °C (–40 °F to +160 °F)
Humidity
+10°Cto+30°C(50°Fto86°F):5%to95%
+3
0 °C to +40 °C (86 °F to 104 °F): 5% to 75%
+40 °C to +50 °C (104 °F to 122 °F): 5% to 45%
Altitude (maximum) 2000 m (6500 feet)
M
aximum input voltage without
damage, Input A and Input B
50 Ω:12V
rms
,35V
peak
for duty factor less than 0.1%
1MΩ, 1X attenuator: 350 V (DC + AC pk) to 440 Hz, falling to 12 V
rms
at 1 M Hz
1MΩ, 10X attenuator: 350 V (DC + AC p k) to 440 Hz, falling to 12 V
rms
at 1 MHz
Maximum input level without
damage, Input C
FCA3003, FCA3103: +34 dBm (12 V
rms
)
FCA3020, FCA3120: +27 dBm (5 V
rms
)
MCA3027, MCA3040: +18 dBm (1.78 V
rms
)
FCA3000, FCA3100, and MCA3000 Series Quick Start U ser Manual 1