Adjustment Procedures
3-2
TDS3000C Series S ervice Manual
Description Example
Minimum
requirements
BNC female-to-female
adapter
Two female BNC connectors Tektronix part number
103-0028-00
Adjustment tool Non conducting shaft with diame-
ter 2.5 mm (0.1 in)
Tektronix part number
003-1433-02
1
The number required is reduced by two if you are adjusting a two-channel oscilloscope.
Overview of the Adjustment Process
The oscilloscope is protected from inadvertently changing factory adjustments
by a rear-panel switch. To perform the factory adjustment, you must push and
hold the switch while you power on the oscilloscope.
Before performing adjustment procedures, you must warm up the oscilloscope at
least ten minutes in an ambient temperature between 20° C and 30° C. Adjust-
ments performed prior to warm-up or outside this temperature range may result
in poor performance. You must also recall the default factory settings before
starting the factory adjustment procedure.
The factory adjustment procedure consists of a series of steps; as you move
through these steps, the oscilloscope display provides instructions that describe
the specific input signal requirements for each step. If the the step passes, the
oscilloscope moves on to the next step. If any step fails, you can repeat the step
or choose to abort the procedure.
NOTE. Do not turn any knobs or push any front-panel buttons other than the
Next Step or Previous Step buttons during the adjustment procedure. Doing so
will cause the oscilloscope to abort the adjustment procedure. The oscilloscope
uses the previous calibration constants if the adjustment procedure is aborted.
The screen does not display the actual oscilloscope settings (such as channel
input impedance, vertical and horizontal settings, and so on) during the
adjustment procedure. The oscilloscope automatically sets the instrument
settings, but these settings may not read out correctly on the display.
During some steps, the instrument may appear to be idle for several minutes
while it is processing information internally. Please have patience.
If all steps in the procedure are completed successfully, a “Pass” message is
displayed and the new calibration constants take affect. If the procedure fails for
any reason, the oscilloscope reverts to the previous calibration settings.

Other manuals for Tektronix TDS3032C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3032C and is the answer not in the manual?

Tektronix TDS3032C Specifications

General IconGeneral
BrandTektronix
ModelTDS3032C
CategoryTest Equipment
LanguageEnglish

Related product manuals