Removal and Installation Procedures
6-- 40
TDS7104 & TDS7054 Service Manual
3. Reinstallation: If your instrument uses Windows 98, reinstall the NLX
battery as follows:
a. Do in reverse step 2a to reinstall the NLX battery.
b. Power on the oscilloscope.
c. After the oscilloscope displays Press <F1> to resume, power off the
instrument and then power it on again.
d. From the Windows operating system select Start, Programs, and then
MS-DOS Prompt.
H If you want your oscilloscope to have access to a network, enter the
following command:
CD ..\TEKTRONIX\CMOS
CMOS --R NET
H If you want your oscilloscope to not have access to a network, but be
able to boot faster, enter the following command:
CD ..\TEKTRONIX\CMOS
CMOS --R NONET
e. Power off your oscilloscope, and then power on your oscilloscope.
4. Reinstallation: If your instrument uses Windows 2000, reinstall the NLX
battery as follows:
a. Do in reverse step 2a to reinstall the NLX battery.
NOTE. Follow the instructions in the Instruction.doc file on the floppy disk.
Viewing this file requires Windows tools. The instrument can not view the file
while booting from the floppy disk.
b. Install the BIOS Update and LAN Enable floppy disk that came with
your Windows 2000 Operating System Restore CD.
c. Remove the hard drive from the instrument.
d. Power on the oscilloscope.
e. If the software suggests that you update your BIOS, let the software
update the BIOS.
f. To install the correct CMOS settings, select either Enable Networking or
Disable Networking. The instrument will boot faster with networking
disabled.