Removal and Installation Procedures
TDS7104 & TDS7054 Service Manual
6-- 41
g. After the software updates your instrument settings, remove the floppy
disk.
h. Power off your oscilloscope.
i. Reinstall the hard drive.
j. Power on your oscilloscope.
1. Locate module to be removed: Locate the NLX Board in the locator diagram
Internal Modules, Figure 6--8, on page 6--19. Additional modules to be
removed:
H Trim (all)
H Bottom cover
H Left and Right covers
2. Remove the NLX Board assembly: See Figure 6--25, on page 6-- 42.
3. Orient the oscilloscope: Set the oscilloscope so its bottom is down on the
work surface and its top panel is facing you.
a. Remove the two T-15 Torx screws that secure the floppy disk drive
assembly into the front chassis.
b. Slide the floppy drive assembly, with cable attached, out toward the rear
of the oscilloscope. Place floppy drive assembly on top of the hard/CD
drive bracket.
c. Remove the five T-15 Torx screws that secure NLX board assembly to
the chassis.
d. Remove the five T-15 Torx screws that secure NLX board assembly to
the rear chassis.
e. Remove the shield between the hard disk drive and the CD drive.
f. Grasp the front edge of the NLX board assembly and pull up on the
assembly to disconnect the Riser Adapter from the Processor board edge
connector.
g. Remove the N LX board assembly from the oscilloscope.
NLX Board

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals