Brief Procedures
TDS7104 & TDS7054 Service Manual
4-9
H Average mode displays an actively acquiring waveform on-screen with
the noise reduced.
H Envelope mode displays an actively acquiring waveform on-screen with
the noise displayed.
H Waveform Database mode (optional) displays an actively acquiring
waveform on-screen that is the accumulation of several acquisitions.
8. Test all channels: Repeat steps 2 through 7 until all four input channels are
verified.
9. Remove the test hookup: Disconnect the BNC cable from the channel input
and the probe compensation output.