EasyManua.ls Logo

Teledyne 10 - Routing the Sample out Lines (Column Module); Pump Controller Connection

Teledyne 10
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BioOptix™ 10
Section 2 Installation
2-4
3. While facing the back of the column module, locate the two
liquid tubing bundles secured to the lower-left inside case
panel. Route these two bundles through the slots at the
edge of the panel (Figure 2-4).
Figure 2-4 Routing the Sample Out lines (column
module)
4. Connect the beige 25-pin connector from the pump module
to the column module connector labeled Pump Controller.
Refer to Figure 2-5.
Figure 2-5 Pump controller connection
Route tubing
through slots
To
Foxy 200
From Pump
Module

Table of Contents