12
Functions
Menu
Setup Notes
(Press F1)
Acquisition mode
See Note A
Sampling
Peak detect
Average
Set to standard sampling mode.
Set to peak detect mode.
Set to average sampling and display the average number of times.
Average number
of times (when
average sampling
is selected)
2~256 Set the average number of times in multiples of 2, i.e. 2, 4,
8, 16, 32, 64, 128, 256. To change the average number of
times, use the jog dial.
Sampling mode
Real time
Equivalent
Set sampling to real time.
Set sampling to equivalent at a time base range of
5ns~100ns/div.
(Press F3)
Display type
See Note B
Vector
Dot
Sampling points are linked for display.
Sampling points are directly displayed.
Duration
Off
2s
5s
Innite
Waveform on the screen is refreshed at normal rate.
Waveform on the screen is maintained for 2 seconds and
then refreshed.
Waveform on the screen is maintained for 5 seconds and
then refreshed.
Waveform on the screen remains on display. New data will
be added continuously.
Format
YT
XT
Operating mode of the oscilloscope.
X-Y is the display mode; A input signal is X signal, B input
signal is Y signal.
(Press F2)
Parameter
Measurement
Customised
All parameters
OFF
The screen displays only user-dened parameters.
All 27 parameters are displayed.
Close the automatic measurement function.
Customised
parameters
Parameter 1
Parameter 2
Parameter 3
Up to 3 parameters can be displayed simultaneously on
screen.
No. of parameters Total 27 Select with the jog dial
Channel
A
B
OFF
Display measurement value of Channel A.
Display measurement value of Channel B.
Close the automatic measurement display.
Indicator
Parameter 1
Parameter 2
Parameter 3
This function offers visual indication of the parameter being
tested.
Frequency
Counter
ON
OFF
Turning the frequency counter on/off
4. Pre-trigger/Delayed Trigger: Data sampled before/after triggering.
The trigger position is typically set at the horizontal center of the screen. In this
case, you are able to view 6 divisions of pre-trigger and delayed information.
Adjusting the horizontal position of the waveform allows you to see more pre-trigger
information. By observing pre-trigger data, you can see the waveform before trigger
occurs. For example, you can detect the glitch that occurs when the circuitry starts.
Observation and analysis of trigger data can help you identify the cause of glitch.
Acquisition Modes, Display and Automatic Measurement
• Press [SCOPE] to access frequently used function options, as shown in the table.