3 Specifications
6
3 Specifications
3.1. Use
Data loggers testo 175 are used to store and to read out
individual readings and measurement series.
With testo 175 measurement values are measured, saved
and transmitted to the PC via USB cable or SD card,
where they can be read out and analysed with the software
testo Comfort Software. With the software the data loggers
can also be individually programmed.
Typical applications
testo 175 T1 and testo 175 T2 are optimally suitable for
temperature measurements in refrigerators, freezers, cold-
storage rooms and cooling shelves.
testo 175 T3 records two temperatures at the same time
and is most suitable e.g. for monitoring the temperature
spreading between feed and return flow in a heating
system.
testo 175 H1 controls the climatic conditions e.g. in
warehouses, offices and in the production area.
3.2. Technical data
testo 175 T1 (0572 1751)
Feature Values
Measurement
parameter
Temperature (°C/°F)
Sensor type NTC temperature sensor internal
Measurement
range
-35 to +55 °C
System accuracy ±0.5 °C (-35 to +55 °C) ± 1 digit
Resolution 0.1 °C
Operating
temperature
-35 to +55 °C
Storage
temperature
-35 to +55 °C
Shop for Data Logging products online at: www.DataLoggerSto re.com 1.888.610.7664