EasyManua.ls Logo

TESTO 425 - Exporting Readings

TESTO 425
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Controls via testo Smart App
27
2
Click on Average of the measured values.
The window for Average of the measured values opens.
3
Enter a value between 2 and 20 measured values.
Changes to the measuring instrument settings in the app are
transferred directly to the measuring instrument. Synchronization with
the app is confirmed on the measuring instrument via “SYNC DONE”.
8.4.4
Setting the Correction Factor
If parts of the cross-sectional area are covered (e.g. by lattice
bars), this
can be corrected via the correction factor.
The correction factor indicates the proportion of free space in the cross-
sectional area.
For example: If 20% of the area is covered, the correction factor must
be set to 0.8 (80% free space).
The correction factor can only be set via the testo Smart App and not
directly on the measuring instrument.
The Settings tab is open.
1
Select Correction factor.
2
Enter the value for the correction factor and save.
The changes are transferred to the measuring instrument and
synchronization with the app is confirmed on the measuring instrument
via “SYNC DONE”.
8.4.5
Setting the area
The cross-sectional area of the measuring point can be set both via the
testo Smart App and directly on the measuring instrument.
The display of this setting on the measuring instrument cannot be
hidden.
The Settings tab is open.
1
Select Area.
1.888.610.7664 sales@GlobalTestSupply.com
www.Testo-Direc t.com

Table of Contents

Related product manuals