10 Technical data for testo 425
33
Technical data for testo 425
• Adjustment conditions for flow probes:
Adjustment in free jet Ø 350 mm, reference pressure 1013 hPa,
based on testo reference Laser Doppler Anemometer (LDA).
• Low flow velocities may result in greater measurem
during humidity and temperature measurement! The probe should
be switched on outside a duct under the following conditions:
Ambient temperature: 20°C
Flow: approx. 0 m/s.
Collapse the telescope after use and do not constrict with the cable.
Start with the telescopic links closest to the handle.
Feature Value
°C, °F
±(0.03 m/s + 4% of measuring value) (0.01 to
20 m/s)
±(0.5 m/s + 5% of measuring value) (20.01 to
30 m/s)
Measuring range 0.01 to 30 m/s
-20 to +50 °C (instrument)
-20 to +70 °C (probe)
0 to 80 %RH / For indoor use only
3 x 1.5 V AA battery
(included in the scope of delivery)
1.888.610.7664 sales@GlobalTestSupply.com
www.Testo-Direc t.com