TECHNICAL SPECIFICATIONS
UNITS: VALUE:
Humidification
Water quality
min. - max. level l
Distilled or fully desalinated
2.2
Water demand to humidify:
from 40 % rel. humidity to 95 % rel. humidity ml 17
Water consumption ml/h 10
Total capacity l 4.7
Measuring range and Setting range* % r.H. 60 ... 95
Control accuracy % r.H. ± 1
GAS ENGINEERING
Carbon dioxide (CO
2
)
Oxygen (O
2
)
Nitrogen (N
2
)
Purity
Inlet pressure
%
bar
99.5
0.8 up to max. 1
Measuring range and Setting range % CO
2
0 ... 20
Control accuracy % CO
2
± 0,1
ELECTRICAL:
Rated voltage
Rated frequency
V
Hz
1/PE AC, 230 or 1/PE AC, 120
50/60 50/60
Radio interference suppression (DIN VDE
0875)
Interference suppression level N
Type of enclosure (DIN 40050) IP 20
Protection class I
Overvoltage category (IEC 1010) II
Degree of soiling (IEC 1010) 2
Rated current A 5,2 or 10,0
Fuse protection:
Fuse
Circuit-break switch
T 16 A slow-blow
G 16
Rated power consumption during
incubation/disinfection
kW 1.2 / 1.2
MISCELLANEOUS
Sound pressure level (DIN 45635 Part 1) dB (A) < 50
max. r.H of the ambiance % r. H 80
Height m NN 2.000
* Under some circumstances, eg. when working with culture vessels without covers, the control
range will be restricted to 90-95% since below this level evaporation of the medium will lead to
an increase of the relative humidity in the chamber.
12. TECHNICAL SPECIFICATIONS
50 079 041 39/45