CO
2
Incubator BB 150 Operating Instructions
42
6. Handling and control
6.10 Resetting the thermal protection
1. The “Thermal Protection Active ” LED illuminates.
2. Turn the device off:
Press the power switch.
All indicators come off.
3. Turn the device back on:
Press the power switch.
NOTE – Thermal protection
When the cause of the failure (e.g. excessive
temperature in the operating room) has been
repaired, the device is set to normal incubation
operation after it has been turned on again. If
the cause of the failure cannot be repaired with
simple measures (e.g. by ventilating the room
or by reducing the temperature in the opera-
ting room), the thermal protection will respond
again immediately; in this case, contact Tech-
nical Service.