10. Maintenance
10.5 Preparing the CO
2
calibration
To determine the exact measured value of the device-integral CO
2
sensor, a CO
2
comparison measurement may be performed
every three months.
If a major deviation is found during this check, a CO
2
calibration
is required.
During this process, the CO
2
control of the device is set to the
value measured during the comparison measurement.
Use a calibrated measuring instrument with an accuracy of
± 0.3 % CO
2
for this test.
Suitable instrument:
Portable IR readout instrument. (Part no. see Section 11,
“Spare parts and accessories”)
The measuring sample is withdrawn through the sealable mea-
surement opening of the glass door. The comparison measu-
rement must be performed when the device is completely stable.
Comparison measurement procedure:
1. Turn device on using power switch.
2. Set CO
2
nominal value and allow device to heat up com-
pletely and to create humidity. This process may take se-
veral hours.
3. Fig. 24: Insert the measuring instrument probe through the
measurement opening [1] into the work space. Wait until
the CO
2
value displayed by the instrument has stabilized.
4. Remove measuring probe, plug measurement opening and
close doors.
5. Calibrate CO
2
control as described in Section 10.6.
Measurement example:
•CO
2
nominal value: 5 %
Measured value: 5.6 %
NOTE – Excessive CO
2
content
Excessive CO
2
content after the calibration can
be reduced by leaving the device doors open
for approx 30 seconds.
Fig. 24
CO
2
calibration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific BB 150 and is the answer not in the manual?

Thermo Scientific BB 150 Specifications

General IconGeneral
BrandThermo Scientific
ModelBB 150
CategoryAccessories
LanguageEnglish

Related product manuals