Thermo Scientific HyPerforma G3Pro Universal Bioreactor Controller User’s Guide | 21
Figure 2.1. TruFlow gas manifold.
The gas supply to the TruFlow should be regulated to between 2 to
2.4 bar (30 and 35 psi) pressure. Since the TruFlow has no regulators,
the regulators must be fitted at the supply tanks, gas lines, or the utility
panel .
Note: Regulators should be installed as close to the TruFlow as
possible as to avoid any flow problems.
2.1.1 Connecting the TruFlow
1. Ensure the TruFlow is placed in a secure location where it cannot tip
or be knocked over. Note: The TruFlow must be placed on top of
the utility tower.
2. Use the respective TruFlow communication cable to connect the
controller to the TruFlow. Refer to Figure 2.2 and Figure 2.3.
Chapter 2 | Gas and liquid handling