Daily Operation
Things to Do After Operating the TSQ System
68 TSQ Series Hardware Manual Thermo Scientific
Go to the next topic, “Cleaning the Ion Sweep Cone and Ion Transfer Tube.
Cleaning the Ion Sweep Cone and Ion Transfer Tube
You must clean the ion sweep cone and the ion transfer tube on a regular basis to prevent
corrosion and to maintain optimum performance of your API source. If you use nonvolatile
buffers in your solvent system or high concentrations of sample, you might need to clean the
ion sweep cone and ion transfer tube more often. See the topic, “Maintaining the Ion Sweep
Cone and Ion Transfer Tubeon page 79.
Go to the next topic, “Emptying the Solvent Waste Bottle.
Emptying the Solvent Waste Bottle
Thermo Fisher Scientific recommends checking the solvent level in the solvent waste bottle
on a daily basis. If necessary, empty the solvent waste bottle. Dispose of the solvent waste in
accordance with local and federal regulations.