Ø1/4" Shutters and Controller Chapter 7: Cleaning and Maintenance
Rev F, January 18, 2023 Page 15
Chapter 7 Cleaning and Maintenance
The SHB025(T) shutter and controller can be cleaned using a soft cloth. If needed,
the cloth can be dampened with some isopropyl alcohol. Avoid using any solvents
on or near the unit as well as water as this may cause the blades to rust.
The SHB025(T) incorporates two shutter blades. These blades are thin (0.003")
and lightweight making them very delicate. Please do not insert any objects into
the shutter where the blades may be hit. This may cause a blade to be slightly bent
hampering performance.
Do not place the shutter head close to any magnetic and ferrous materials. It may
affect the performance of the coils within the shutter. Maintain a distance of at least
1/8" from any ferrous materials. While both sides are susceptible, the front face is
more susceptible to interference.