© 2017 Thorlabs GmbH24
TXP5004
7.2 Certifications and Compliances
EC Declaration of
Conformity - EMC
Meets intent of Directive 2004/108/EC
1
for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EMC requirements for Class A electrical equipment for measurement,
control and laboratory use, including Class A Radiated and Conducted
Emissions
2,3,4
and Immunity
2,3,5
.
Electrostatic Discharge Immunity (Performance Criterion C)
Radiated RF Electromagnetic Field Immunity
(Performance Criterion A)
Electrical Fast Transient / Burst Immunity (Performance Criterion C)
Power Line Surge Immunity (Performance Criterion C)
Conducted RF Immunity (Performance Criterion A)
Voltage Dips and Interruptions Immunity (Performance Criterion C)
AC Power Line Harmonic Emissions
Voltage Fluctuations and Flicker
Australia / New Zeeland
Declaration of
Conformity - EMC
Complies with the Radiocommunications Act and demonstrated per EMC Emission Standard
2,3,4
Industrial, Scientific and Medical Equipment: 1992
Emissions comply with the Class A Limits of FCC Code of Federal Regulations 47, Part 15,
Subpart B
2,3,4
.
EC Declaration of
Conformity - Low
Voltage
Compliance was demonstrated to the following specification as listed in the Official Journal of the
European Communities:
Low Voltage Directive 2006/95/EC
6
Safety Requirements for Electrical Equipment for Measurement, Control
and Laboratory Use
U.S. Nationally
Recognized Testing
Laboratory Listing
UL3111-1
ANSI/ISA
S82.01:1994
Standard for Electrical Measuring and Test Equipment
Safety Standard for Electrical and Electronic Test, Measuring,
Controlling and Related Equipment
CAN/CSA C22.2 No.
61010-1 3
nd
ed.
Safety Requirements for Electrical Equipment for Measurement, Control
and Laboratory Use
Safety Requirements for Electrical Equipment for Measurement, Control
and Laboratory Use
Class I equipment according to IEC 61010-1, Annex H, grounded product
1
Replaces 89/336/EEC.
2
Compliance demonstrated using high-quality shielded interface cables shorter than or equal to 3 meters.
3
Compliance demonstrated with various TX Series light source and electrical modules installed
4
Emissions, which exceed the levels required by these standards, may occur when this equipment is connected to a test object.
5
Minimum Immunity Test Requirement
6
Replaces Low Voltage Directive 73/23/EEC, amended by93/68/EEC