EasyManua.ls Logo

Topcon RL-VH3A - Vertical Calibration Check Details (RL-VH3 G;A)

Topcon RL-VH3A
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
Vertical Calibration (Upward)
Perform the following check after completing "Horizontal Calibration" on the previous page.
(1) Checking RL-VH3G/A
1 Set up the instrument half way between 2
walls a minimum of 40m away from each
other. (The instrument can be facing
either direction X or Y. And no tripod is
used.)
2 Turn the power switch on.
3 Place a piece of paper on each wall (A and B).
Mark the horizontal laser positions (Ma and Mb) on each wall using target.
4 Turn the power switch off. Position the
instrument for vertical operation (see
instruction on page 19 ) with the bottom
side directly facing wall A (see illustra-
tion). Make sure the unit is level by check-
ing the circular level vial. Use the leveling
screw to adjust if necessary.
5 Turn the power switch on. (Laser beam
should be in scanning mode.)
6 Mark where the split beam is emitted
from the bottom and top of the instrument
strikes wall A and B. Measure the distance between two marks on each wall (dHa and
dHb).
about 20 m
(66ft)
about 20 m
(66ft)
Wall
A
Wall
B
about 40 m (130ft)
Wall
A
Wall
B
about 1m

Other manuals for Topcon RL-VH3A

Related product manuals