EasyManua.ls Logo

TPS miniCHEM-pH - Page 13

TPS miniCHEM-pH
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 13
3.5.3 Adjusting the Hysteresis
The hysteresis is the pH range over which the alarm/control output remains
switched on, once the trip point has been exceeded.
For example, if the limit is set to pH7.00, as a “too low” trip point, with a
hysteresis of ±0.10pH, the alarm/control output is switched ON when the reading
goes to pH6.90 (i.e. 7.00 - 0.10). The output is not switched OFF until the reading
goes back up to pH7.10 (i.e. 7.00 + 0.10).
The hysteresis stops the relay “chattering” around the set point, by providing a
buffer between the points at which the output is switched ON and OFF. This
feature can significantly increase the life of pumps and solenoid valves.
The factory-set hysteresis is approximately ±0.10 pH, as in the example above.
The hysteresis can be increased to as high as ±0.40 pH with the HYSTERESIS
trimmer in the terminal area. Turning the HYSTERESIS clockwise increases the
hysteresis and turning it anti-clockwise decreases the hysteresis.
TPS DOES NOT recommend that this setting be altered, unless it is absolutely
necessary.
The HYSTERESIS trimmer should only be adjusted a little at time. The process
being controlled should then be closely monitored over a time to ensure that the
desired control is being achieved.

Related product manuals