Refilling Reference Electrode
The Reference Electrode cap needs to be removed prior to analysis and
replaced when the electrode is going to be stored. It is best practice not to let
the tip of the Reference Electrode dry out. To ensure this, always add a few
drops of deionised water into the cap before replacing on the electrode.
The Reference Electrode must contain Reference Electrode Fill Solution to
operate. Ensure the Reference Electrode is held vertically with the connector
upwards and determine that there is enough solution in the Reference
Electrode.
Fill Solution at max Fill solution level too low
The outer tube needs to be at least 1/3rd full. To top up the Reference
Electrode Fill Solution:
1. Place the reference electrode horizontal with the hole facing up
2. Slide the band down from the hole
3. Use the pipette to drop enough solution into the Reference Electrode to fill
it up.
4. Once topped up, replace the band to prevent the electrode fill solution
escaping.
Note: Only use the supplied Reference Electrode Fill Solution.
Metalyser® Field Pro HM3000 Instruction Manual v 5.0
www.trace2o.com
14

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Trace2O METALYSER FIELD PRO HM3000 and is the answer not in the manual?

Trace2O METALYSER FIELD PRO HM3000 Specifications

General IconGeneral
BrandTrace2O
ModelMETALYSER FIELD PRO HM3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals