To reset the meter, hold the TARE ^ RESET key depressed while you press and
release the > MENU key.
To reset a Peak, Valley or First Peak, hold the TARE ^ RESET key depressed
while you press and release the > PEAK key.
To reset latched Alarms, hold the TARE ^ RESET key depressed while you press
and release the < ALARMS key.
The TARE ^ RESET key is also used in the Menu mode and in the RECALL
mode to increment the flashing selection.
< ALARMS From the RUN mode, press < ALARMS to change to the Alarms mode, which
allows viewing and changing setpoints from the keypad. Press once to view
Setpoint 1. The Alarm 1 indicator (amber LED) will light. Press again to view
Setpoint 2. The Alarm 2 indicator (red LED) will light. Press again to exit the
Alarms mode and return to the RUN mode. While the Alarm 1 indicator is lit,
you can change Setpoint 1 by pressing > PEAK to select a digit, which will flash,
then pressing TARE ^ RESET to increment that digit. Same for Setpoint 2.
In the Menu mode, the > ALARMS key steps backward through the menu. In the
RECALL mode, it is used to decrement Memory location #’s from 15999 to 0.
TEDS ERROR MESSAGES
If there is a problem with the TEDS connection between the meter and TEDS transducer,
one of the following error messages is displayed at power on:
Err 1 - TEDS data line shorted.
Err 2 - No Presence pulse from TEDS.
Err 3 - Improper Presence pulse width.
Err 4 - Error in TEDS family code in ROM.
Err 5 - TEDS checksum error in the first or second 32 bytes.
Correct the problem and then press the Menu key.