EasyManuals Logo

Transducer Techniques SSI User Manual

Transducer Techniques SSI
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
- 41 -
16. SPECIFICATIONS
Display
Type ........................................................ Reflective LCD, 2.56” x 1.73” (65 x 44 mm) window
Main numeric display ........................................ 6 seven-segment digits, 0.57” (14.5 mm) high
Other displayed items ............................. 5-digit recall number, choice of 18 units of measure,
battery life indicator, operating mode indicator
Overload indication ............................................................ Display flashes at 120% of HiRead
Response Rates
Normal data sampling rate .......... 15,360 samples/second (setting for 60 Hz noise rejection)
12,800 samples/second (setting for 50 Hz noise rejection)
Normal conversion rate ............................................................................. Every 256 samples
60/second (setting for 60 Hz noise rejection)
50/second (setting for 50 Hz noise rejection)
Fast conversion rates .............................................. Every 2 to 128 samples (programmable)
Relay response rate ............................................................ Same as Normal conversion rate
Display update rate ............................................................................... Every 16 conversions
3.750/second (setting for 60 Hz noise rejection)
3.125/second (setting for 50 Hz noise rejection)
Peak & valley capture rate (selectable) ............ Every 2, 4, 8, 16, 32, 64, 128 or 256 samples
Peak & valley display update rate ................. 3.750/second (setting for 60 Hz noise rejection)
...................................................................... 3.125/second (setting for 50 Hz noise rejection)
Analog-to-Digital Conversion
Recommended load ................................................................................ 120-10K ohm bridge
Bridge excitation voltage .................................................................................................. 3.0V
Provision for variations in bridge excitation ............................................ Ratiometric operation
Signal span, max ........................................................................................................... 15 mV
Transducer sensitivity range ................................................................................... 1-5 mV / V
A-to-D converter resolution ................................................................. 16 bits (65,536 counts)
Signal resolution .................................................................................................. 0.5 µV/count
Accuracy at 25°C ......................................................................................................... ±1.5 µV
Span Tempco ....................................................................................... 0.0015% of reading/°C
Zero Tempco ............................................................................................................ 0.2 µV/°C
NMR with no filtering ...................................................................................................... 90 dB
Signal filtering (selectable) .................................. 1) No filtering: 50 or 60 conversions/second
2) Batch average of 16 conversions
3) Moving average of conversions
Power Requirements
Internal battery ................................................................................. Rechargeable lithium ion
Battery capacity (nominal) .............................................................................. 2400 mAh, 3.7V
RUN time on single battery charge .................................. Up to 65 hours with 350 ohm bridge
Battery charger .................................................................................. Wall plug AC power unit
Battery charger input ................................................................ 100-240V ac, 50/60 Hz, 0.15A
Battery charger output ........................................................................................ 5.0V dc, 1.0A
Battery charging time, max ........................................................................................... 8 hours

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Transducer Techniques SSI and is the answer not in the manual?

Transducer Techniques SSI Specifications

General IconGeneral
BrandTransducer Techniques
ModelSSI
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals