B-2 Model 3775 Condensation Particle Counter
RSF Read the aerosol flow rate in cc/min
Returns: A floating point number from 0.0 to 9999.9
Example: 300.0
RIF Read the inlet flow rate setting in liters per minute
Returns: A floating point number from 0.0 to 9999.9
Example: 0.3
RTS Read the saturator temperature in degrees Celsius
Returns: A floating point number from 0.0 to 50.0
Example: 39.0
RTC Read the condenser temperature in degrees Celsius
Returns: A floating point number from 0.0 to 50.0
Example: 14.0
RTO Read the optics temperature in degrees Celsius
Returns: A floating point number from 0.0 to 50.0
Example: 40.0
RTA Read the cabinet temperature in degrees Celsius
Returns A floating point number from 0.0 to 50.0
Example 23.8
RCT Read the current time
Returns: Www Mmm dd hh:mm:ss yyyy
where
Www is the weekday
Mmm is the month in letters
dd is the day of the month
hh:mm:ss is the time
yyyy is the year
Example: Mon Jun 11 11:05:08 2006