4-10 Model 3775 Condensation Particle Counter
Option Concentration Range for Analog Output 0–10 V Relation
OFF 0 V independent of concentration -
1E+1 0 to 10 particles/cm
3
linear
1E+2 0 to 100 particles/cm
3
linear
1E+3 0 to 1,000 particles/cm
3
linear
1E+4 0 to 10,000 particles/cm
3
linear
1E+5 0 to 100,000 particles/cm
3
linear
1E+6 0 to 1,000,000 particles/cm
3
linear
1E+7 0 to 10,000,000 particles/cm
3
linear
LOG 10 V = 10,000,000 particles/cm
3
9 V = 1,000,000
8 V = 100,000
7 V = 10,000
6 V = 1,000
5 V = 100
4 V = 10
3 V = 1
2 V = 0.1
1 V = 0.01
0 V = <0.01
log
Drain
For specific instructions on draining butanol, refer to the section
Draining Butanol from the Butanol Reservoir” in Chapter 8.
During draining, the Auto-Fill mode and the high vacuum pump
operation are automatically turned off. When draining is stopped,
the pump restarts but the Auto-Fill must be turned on again by
selecting this option from the User Settings menu. Whenever the
instrument is turned on, the Auto-Fill is activated. Refer to the
previous caution note.
Graph Options
Refer to Figure 4-2 depicting the graph while reading this section.
Figure 4-7 shows the options possible when Graph Options is
selected from the User Settings Menu. A description of these
options is provided below:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 3775 and is the answer not in the manual?

TSI Instruments 3775 Specifications

General IconGeneral
BrandTSI Instruments
Model3775
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals