EasyManua.ls Logo

TSI Instruments 3775 - Bypass;Makeup Air Flow Adjustment; Installing a Saturator Wick

TSI Instruments 3775
133 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Service 8-15
Figure 8-10
Installing a Saturator Wick
Bypass/Makeup Air Flow Adjustment
Refer to the flow schematic in Figure 5-1.
High inlet flow is adjustable using the bypass/makeup air variable
orifice. The bypass/makeup air is nominally 1.2 L/min. Together
with nominally 0.3 L/min aerosol flow, 1.5 L/min inlet sample flow
is produced. Aerosol flow is not adjustable because it is controlled
by the critical orifice.
To make adjustments to the bypass/makeup air flow, you must
have a suitable flowmeter as a reference. Choose an external
flowmeter that has a low pressure drop and measure the actual
volumetric flow e.g., a bubble meter or a TSI flowmeter. If you use a
mass flowmeter (one referenced to standard conditions), convert the
standard flow to actual (volumetric) flow with the following
equation:
××=
P
T
sscmscm
35.101
293
)/( flow std)/(flow Actual
33
where T = °K
P = kPa

Table of Contents

Related product manuals